天天干亚洲_99久久免费国_亚洲国产精品成人久久久软件_亚洲成av人**亚洲成av**_精品中文字幕在线AⅤ_男女真实毛片视频图片

English | 中文版 | 手機(jī)版 企業(yè)登錄 | 個(gè)人登錄 | 郵件訂閱
當(dāng)前位置 > 首頁 > 技術(shù)文章 > 寨卡病毒核酸檢測試劑盒英文使用說明書

寨卡病毒核酸檢測試劑盒英文使用說明書

瀏覽次數(shù):6395 發(fā)布日期:2016-9-28  來源:本站 僅供參考,謝絕轉(zhuǎn)載,否則責(zé)任自負(fù)

寨卡病毒核酸檢測試劑盒(德國Genekam原裝進(jìn)口)英文使用說明書

Lot-No.
Ref. FR340 Expiry time: 1 year

100 Tests (Ready to use PCR kit)  

Zika Virus    (Real time) DOUBLE CHECK                                                    

Only for in vitro use -

only for research use

-To be used by a technical person-

Principle and use

This amplification kit has been manufactured by Genekam Biotechnology AG, Germany to detect Zika-Virus (Double check. This is an absolute quantification assay. It tests twice the samples.

Real time PCR is based on fluorogenic dyes. In our kit we use 2 dyes (2 probes), first Probe they are 6-Carboxy tetramethyl rhodamine (quencher) and Carboxy-fluorescein (reporter; called also FAM-Channel). Second Probe is labelled with HEX reporter (Filter for is VIC or in some machines, it is HEX)on one side and on the other side 6-Carboxy tetramethyl rhodamine (quencher). Up to 40 Ct for each dye should be taken positive. Value between 40-45 Ct for each dye should be taken as marginal positive (doubtful).
This kit needs RNA which can be isolated from blood, serum, oral samples, liver samples, urine samples, spleen, brain tissue, vaccine, seminal samples, lymph nodes, cell cultures, tissue and any body fluid. Kindly use good methods to isolate the RNA. Kindly take common safety laboratory precautions during working  as Zika virus is highly infectious. Please use gloves during work and masks. Proceed clean and carefully otherwise you may cause contamination problems. Do not touch other objects like pens, chairs etc. during Part 1.

IMPORTANT: we added cotton or sponge in the lid of container of the kit, to avoid damage during transportation. Please remove this cotton or sponge from the lid of each container before storage.

Composition:

It contains the following: (Warning: THAW THE TUBES SLOWLY: NEVER THAW IN HEATING BLOCK OR WITH HEAT FROM HAND):

• Tube A (2 tubes)

• Tube B (2 tubes)
• Tube Y (1 tube)

• Positive (+Ve) control (D1): to be stored at -20°C (1 tube): It must be used with pipette tips with filter or it should be used with separate pipettor.

• Negative (-Ve) Control (tube D2) (1 tube)

Please check them before you start. Please keep all tube away from light.

Equipments needed:

• Realtime PCR thermocycler

• Laboratory centrifuge

• microtubes (0.2ml)

• sterile Pipette-tips with and without filter (20µl, 5µl & 1µl)

• Pipettes (quality pipettes)

• Paper

• Pen

• Microtube

• Ice

• Vortexer
• 96 well PCR plate

Procedure:

ONCE AGAIN:

VERY IMPORTANT !  PLEASE USE GLOVES ! DON’T TOUCH ANY OTHER OBJECTS, OTHERWISE THERE MAY BE RNASE CONTERMINATION DURING THIS PART.

STEP A

1. Kindly thaw one tube of: A, B, Y, D1 and D2. After thawing, kindly put the tubes at 4°C (as it is better). If the kit is not in use, store them at -20°C.

2. Mark your microtubes with a sample number,  +ve Control and –ve Control. You can use 96 well microplate instead of tubes.

3. Add 7µl of tube A to each tube.

4. Add 10µl of B to each micro tube. Avoid to touch the wall of the microtubes.


5. Add 1µl of Y to each tube (avoid to touch the wall of the microtubes).

TIP: Add 7µl A + 10µl B + 1µl Y = 18µl per reaction. In case you want to run 10 reactions i.e. you need total 180µl, therefore you should mix 70µl of A + 100µl of B + 10µl of Y = 180µl from which you can take 18µl and add to each tube. This way you save time and hardware.

6. Add 2µl of your RNA with sterile pipette-tip with filter to each micro tube according to your label except +Ve and -Ve (Avoid touching the wall).  Use every time a new pipette tip (for each sample) !

7. Use new pipette tip with filter. Add 2µl of tube D1. This is the positive control supplied with our kit. Mix it.

8. Use a new pipette tip. Add 2µl of –Ve (tube D2) to –Ve Control (don’t touch the wall).Mix it.

9. Centrifuge all tubes for 20 sec. for 8000 rpm (this is not necessary but it is better).
10. Run the program of your thermocycler as followings: Kindly check whether you have added everything correctly as the level of the volume of each micro tube must be almost the same.

You must use quencher and reporter dye to setup your software (see FAQ) and run the following program:

1. 60 minutes at 42°C
10 minutes at 70°C

2. 15 seconds at 95°C   x 45 cycles
60 seconds at 60°C

Before you start the PCR program, kindly check whether tubes are closed properly. Microtubes must be in contact with metal block (important!). There should be no air or lose contact with metal block of thermocycler.

11. After step 10 is finished take out the microtubes.

STEP B

Once the program will be finished one can see the graphics. The negative control should run along with the bottom and positive control must give a curve in the software graphics. Use your software to analyse the results.

Results: Usually there will be samples with two curves i.e. they are positive. There will be samples with one curve i.e they are positive. Samples without any curve are negative. The samples with two curves should have two Ct values i.e. they are double positive. The samples with one curve has only one Ct value i.e. positive samples. The samples without any curve will have no Ct values i.e. they are negative.

If you should find any mistakes, please let us know. Thank you.

Suggestion:
This manual has been written specifically for beginners, hence persons with experience in PCR must use their experience to keep each step as small as possible e.g. you should calculate the amount of the needed chemicals, before starting with testing.

 

Last update: 15--12-2016

v1.0

Genekam Biotechnology AG
Duissernstr. 65a
47058 Duisburg

Germany

Tel. (+49) 203 / 555858-31,-32,-33

Fax (+49) 203 / 35 82 99

anfrage@genekam.de

http://www.genekam.de


FAQ:

1) Q: I cannot find quencher and reporter dye in my software:

A: Many software has got the words:  FAM (as reporter) and TAM (as quencher). Therefore select both in your software.

If your machines has only one word (for some machines only use the word FAM) you should select this one.  For HEX, you can use HEX or VIC or JOE in your Software.

2) Precautions during the pipetting: Positive control should be pipetted with a separate pipettor and it should not be used for pipetting unknown samples or one should use pipette tips with filte

發(fā)布者:深圳市科潤達(dá)生物工程有限公司
聯(lián)系電話:0755-26814430
E-mail:kerunda@163.com

用戶名: 密碼: 匿名 快速注冊 忘記密碼
評(píng)論只代表網(wǎng)友觀點(diǎn),不代表本站觀點(diǎn)。 請輸入驗(yàn)證碼: 8795
Copyright(C) 1998-2025 生物器材網(wǎng) 電話:021-64166852;13621656896 E-mail:info@bio-equip.com
主站蜘蛛池模板: 精品国产一区二区三区成人影院 | 国产自慰网站在线观看 | 国产三级久久三级久久 | 久热网站 | 欧美91在线 | 国产v日韩v欧美v在线观看 | "a级毛片 黄" | 国产黄色小视频 | 久久大屁股少妇 | 在线播放一级片 | 天堂久久一区 | 国产成人精选在线观看不卡 | 无码精品国应Aⅴ左线 | 宅男在线观看免费高清网站 | 国产精品一区二区二 | 精品国产一区二区三区四区在线看 | 色婷婷久久综合中文久久一本 | 亚洲欧美日韩中文字幕在线一区 | 亚洲一级影片在线观看 | 夜爽8888视频在线观看 | 精品色999| av免费观看网页 | 亚洲精品97 | 久久一区三区 | 亚洲人成网站777色婷婷 | 黄瓜福利视频在线观看 | 国产精品一级a级理论片在线观看 | 一级黄a| 久久久久久伊人高潮影院 | 久久夜夜躁躁精品 | 国产欧美日韩一区综合在线视频免费 | 不卡视频一区二区三区 | 看美女pp日出水了免费视频 | 亚洲成人免费观看 | 国产在线观看91精品不卡 | 国产福利姬喷水福利在线观看 | 看a黄视频 | 午夜国产情侣在线播放 | 国产第一页视频在线播放 | 又大又粗又爽αa级毛片 | 97超碰免费人人看人人 |